介绍了基于光腔衰荡光谱法测定气体中的微量水的方法原理,该方法适用于含量为0.2×10-9~20×10-6 mol/mol的所有气体中水分的测定。建立了光谱分析法的原始数学模型,对测定方法的不确定度进行了评估,结果表明该方法的扩展不确定度为(8~20)×10-9 mol/mol(k=2,p=95%)。
Abstract
A new method for determining the moisture in gases by the cavity ring down spectroscopy is available to measure the concentration between 0.2×10-9 mol/mol and 20×10-6 mol/mol.A mathematical model is established for the uncertainty evaluation,and the results of the expanded uncertaunties are in the range of(8~20)×10-9 mol/mol (k=2, p=95%).
关键词
计量学 /
光腔衰荡光谱法 /
气体湿度 /
不确定度
Key words
Metrology /
Cavity ring down spectroscopy /
Gas moisture /
Uncertainty
{{custom_sec.title}}
{{custom_sec.title}}
{{custom_sec.content}}
参考文献
[1]Brewer P J,MiltonM J T,HarrisP A,et al.International comparability in measurements of trace water vapour[R].NPL Report AS59, 2011.
[2]Yan W B,Chen Y,Chen H, et al.Development and applications of continuous-wave cavity ring-down spectroscopy[J].Int J Thermophys,2008, 29(5): 1567-1577.
[3]JCGM100—2008.Guide to the Expression of Uncertainty in Measurement[S].