Abstract:Temperature standard instruments are studied for the temperature sensor pair of heat meter. Two grade B industrial platinum resistance thermometers (IPRTs) are tested. The experiments showed that the stability of high precise IPRT is 3 mk in the temperature range from 0 ℃ to 90 ℃ per year.The resistance drifts of the IPRTs were eliminated by using W value calculation and the accuracy of the calibration is improved.
邱萍,谢立,王池. 检定热量表配对温度传感器的温度标准实验研究[J]. 计量学报, 2011, 32(z1): 74-76.
QIU Ping,XIE Li,WANG chi. Research on the Temperature Standard for the Thermometer Sensor Pair of Heat Meter. Acta Metrologica Sinica, 2011, 32(z1): 74-76.