按照砝码质量测量时对空气密度测量的需求,分析了测量空气密度的公式法,研制了用于高准确度砝码质量测量的空气密度测量装置。根据CIPM-2007公式,该装置集成了温湿度、压力和CO2含量的测量,实现了对空气密度的自动实时测量。实验结果表明,当CO2含量达到937.4×10-6时,与采用近似公式进行测量的数据的相对误差达到0.022%,验证了增加CO2含量测量对提高空气密度测量准确度的必要性。
Abstract
To determine air density for high accuracy measurement of the weights, the method using the recommended equation is studied. According to the CIPM-2007 equation, an equipment of air density measurement that for high accuracy measurement of the weights is developed. Measurement of temperature, humidity, pressure and CO2 are integrated on it to measure air density automatically. Comparing with the results using approximation formula for air density, the relative error of 0.022% is achieved when CO2 is 937.4×10-6. The experimental results prove that additional measurement of CO2 can improve accuracy of air density measurement.
关键词
计量学 /
空气密度 /
质量 /
砝码 /
空气浮力修正 /
CIPM-2007公式
Key words
Metrology /
Air density /
Mass /
Weight /
Air buoyancy correction /
CIPM-2007 equation
{{custom_sec.title}}
{{custom_sec.title}}
{{custom_sec.content}}
参考文献
[1]罗志勇,刘子勇,刘吉萍,等.便携式空气密度精密测量仪的研制[J].计量学报,2001,22(1):52-56.
[2]JJG 99—2006,砝码检定规程[S].
[3]OIML R 111-1:2004, Weights of classes E1, E2, F1, F2, M1, M1-2, M2, M2-3, and M3 Metrological and technical requirements [S].
[4]Picard A, Davis R S, Glser M, et al. Revised formula for the density of moist air (CIPM-2007)[J]. Metrologia,2008, 45:149-155.
[5]Harveya A H, Lemmon E W. Correlation for the second virial coefficient of water[J]. J Phys Chem Ref Data,2004, 33(1):369-376.
[6]Harvey A H, Huang P H.First-Principles Calculation of the Air-Water Second Virial Coefficient[J]. International Journal of Thermophysics, 2007,28(2): 556-565.